Systec Autoclave 75 L capacity

Product image: Vertical Stainless Steel Autoclave VX-75, 75 l Chamber Volume
Back to overview
Item no. 18334

Vertical Stainless Steel Autoclave VX-75, 75 l Chamber Volume

For demanding sterilization processes with all options for process optimization

Product information

Fully Automatic Sterilization for Laboratories with 12 Programs

The VX-75 stand-alone autoclave is suitable for demanding sterilization processes in research laboratories, biotechnology and microbiology. With a usable volume of 75 liters, it reliably sterilizes instruments, liquids and waste in disposal bags. The vertical stand-alone design allows for space-saving installation and convenient top loading.

The fully automatic microprocessor control offers 12 individually programmable sterilization programs for solids, liquids and disposal sterilization. Operating pressure and temperature reach up to 4 bar at 140°C. An integrated separate steam generator, electronic pressure sensor and a validation port for external temperature sensors ensure precise, GMP-compliant processes with an RS 232 interface.

The VX-75 combines process reliability and flexibility for demanding sterilization requirements in the laboratory. You can find further laboratory solutions in our sections Refrigeration Technology and Cryotechnology. Request an individual quote now!

Frequently Asked Questions About This Model

What is the VX-75 used for?

The VX-75 is used for sterilizing instruments, liquids and waste in disposal bags in laboratories in biotechnology, microbiology and pharmaceuticals.

What temperature and pressure are achieved?

The autoclave operates with temperature and steam pressure up to 140°C at 4 bar, controlled by an electronic pressure sensor.

Which laboratories is the VX-75 suitable for?

Thanks to the 12 programmable sterilization programs and validation options, the VX-75 is suitable for research laboratories, biotech companies and facilities with GMP-compliant documentation requirements.

How does the VX-75 differ from other stand-alone autoclaves?

Compared to the simpler VE series, the VX-75 offers comprehensive options for process optimization, such as a validation port, RS232 interface and preparation for rapid cooling, vacuum and drying.

What accessories or maintenance are required?

The VX-75 is prepared for the subsequent installation of a printer, rapid cooling, vacuum equipment and drying; low water level protection and overtemperature protection are integrated as standard.

Technical data & description

Chamber dimensions 400 mm diameter, 600 mm depth
Usable chamber volume total/nominal 85/75 l

Standard Equipment:
• Fully automatic microprocessor control with 12 programs:
Individual program parameters such as temperature and time can be individually selected and stored for each program.
Program 1-3: Fast program for solids, instruments, equipment, plastic items with rapid steam discharge
Program 4-5: For waste in destruction bags with pulsating heating phase for improved venting of destruction bags
Program 6-7: For destruction sterilization of liquids with slow steam discharge. (Or fast recooling as option)
Program 8-11: For liquids with subsequent controlled steam discharge and self-cooling (Or fast recooling as option)
Program 12: Cleaning program with rapid steam discharge

Validation, Documentation:
A 1/2" validation port is available for the installation of external temperature sensors for validation purposes. For documentation of program sequences, various systems are optionally available ranging from built-in printer to PC software via RS 232 interface or separate data loggers for GMP-compliant documentation.

• Integrated separate steam generator (no heating elements in the sterilization chamber)
• Novel safety ring closure system with lip seal
• Temperature and steam pressure up to 140°C / 4 bar
• Thermal barrier, pressure and media temperature controlled temperature-dependent door locking in accordance with DIN and TRB
• Flexible PT-100 temperature sensor for reference vessel
• Media temperature sterilization time triggering
• Electronic pressure transducer for pressure and temperature control
• Autofill, automatic deionized water feed for steam generation
• Special program for destruction sterilization with pulsating heating phase for improved venting of destruction bags
• Start by Clock / Timer for programmable start time
• Exhaust steam condensation water-cooled thermostatically controlled
• Collector connection for all outlet lines
• Drain valve
• Base plate for placement of vessels, baskets, etc.
• Validation port
• RS 232 interface for external data transfer
• Prepared for printer connection (for subsequent installation)
• Prepared for fast recooling (for subsequent installation)
• Prepared for vacuum device (for subsequent installation)
• Prepared for drying (for subsequent installation)
• Water shortage safety device
• Over-temperature protection

External dimensions: 550 x 780 x 995 mm (W x D x H)
Net weight 130 kg
Heating capacity: 9.0 kW
Voltage: 380 - 400V, 50/60 Hz, 16A
Permissible operating pressure: 4 bar absolute
Permissible operating temperature: 140°C

Site-supplied Media Connections:
Electrical connection 380-400V, 50/60 Hz, 16A
Demineralized water: Shut-off valve 3/4" for hose connection (only for permanent installation, water can also be supplied from a storage tank)
Raw water: Shut-off valve 3/4" for hose connection (for exhaust steam condensation, device can also be operated without this)
Drain: Hose connection 3/4" max. 500 mm from floor or floor drain

Warranty: One year from delivery for material and manufacturing defects
3 years on heating elements
5 years on pressure vessel and closure mechanisms
Large product image