Systec Autoclave 120 L capacity

Product image: Vertical Stand Autoclave VX-120, 120 l Chamber Volume
Back to overview
Item no. 18336

Vertical Stand Autoclave VX-120, 120 l Chamber Volume

for demanding sterilization processes with all options for process optimization

Product information

Fully Automatic Floor-Standing Autoclave for Validated Sterilization Processes

The Floor-Standing Autoclave VX-120 with 120 liter chamber volume is suitable for demanding sterilization processes in microbiology, biotechnology, and pharmaceutical laboratories. With 12 individually programmable cycles, the device covers the sterilization of solids, instruments, liquids, and waste in disposal bags. Ideal for facilities with high documentation and validation requirements.

The fully automatic microprocessor control operates with temperature and steam pressure up to 140°C / 4 bar. A separate steam generator, electronic pressure transducer, and a 1/2" validation port enable GMP-compliant documentation. The heating power is 9.0 kW at 380-400V connection, and the device is prepared for rapid cooling, vacuum, and drying.

With extensive process optimization and validation options, the VX-120 offers a powerful solution for professional sterilization requirements. Further information on suitable systems can be found in the Refrigeration Technology and Cryotechnology sections. Request an individual quote now!

Frequently Asked Questions About This Model

What is the VX-120 used for?

The VX-120 is used for sterilizing solids, instruments, liquids, and waste in disposal bags – suitable for demanding sterilization processes in laboratory and research settings.

Which technical specifications are decisive?

The device operates with a chamber volume of 120 liters (nominal), achieves temperatures and steam pressure up to 140°C / 4 bar, and features a heating power of 9.0 kW.

Which laboratories is the VX-120 suitable for?

Due to the 12 programs, validation options, and GMP-compliant documentation options, the device is suitable for laboratories in microbiology, biotechnology, medicine, and pharmaceuticals with high demands on process control.

How does the VX-120 differ from simpler autoclaves?

Unlike simpler models, the VX-120 offers all options for process optimization including a validation port, electronic pressure transducer, and preparation for rapid cooling, vacuum, and drying.

What maintenance or accessories are required?

The device is prepared for the subsequent installation of a printer, rapid cooling, vacuum equipment, and drying. Low-water protection and overtemperature protection for operational safety are included as standard.

Technical data & description

Chamber dimensions 500 mm diameter, 600 mm depth
Total/nominal chamber volume 137/120 l

Standard equipment:
• Fully automatic microprocessor control with 12 programs:
Individual program parameters such as temperature and time can be freely selected and stored for each program.
Program 1-3: Fast program for solids, instruments, equipment, plastic articles with rapid steam exhaust
Program 4-5: For waste in destruction bags with pulsating heating phase for improved venting of destruction bags
Program 6-7: For destruction sterilization of liquids with slow steam exhaust (or rapid recooling as option)
Program 8-11: For liquids with subsequent controlled steam exhaust and self-cooling (or rapid recooling as option)
Program 12: Cleaning program with rapid steam exhaust

Validation, Documentation:
A validation port 1/2" is available for introducing external temperature probes for validation purposes. For documentation of program sequences, various optional systems are available ranging from built-in printer to PC software via RS 232 interface or separate data loggers for GMP-compliant documentation.

• Integrated separate steam generator (no heating elements in sterilization chamber)
• Novel safety ring closure system with lip seal
• Temperature and steam pressure up to 140°C / 4 bar
• Thermal barrier, pressure and media temperature controlled temperature-dependent door lock according to DIN and TRB
• Flexible PT-100 temperature probe for reference vessel
• Media temperature sterilization time release
• Electronic pressure transducer for pressure and temperature control
• Autofill, automatic deionized water feed for steam generation
• Special program for destruction sterilization with pulsating heating phase for improved venting of destruction bags
• Start by Clock / Timer for programmable start time
• Exhaust steam condensation water-cooled thermostatically controlled
• Collective connection for all outlet lines
• Drain valve
• Base plate for placement of vessels, baskets, etc.
• Validation port
• RS 232 interface for external data transmission
• Prepared for printer connection (for subsequent installation)
• Prepared for rapid recooling (for subsequent installation)
• Prepared for vacuum device (for subsequent installation)
• Prepared for drying (for subsequent installation)
• Low water safety device
• Overtemperature protection

External dimensions: 650 x 900 x 985 mm (W x D x H)
Net weight 180 kg
Heating capacity: 9.0 kW
Voltage: 380 - 400V, 50/60 Hz, 16A
Permissible operating pressure: 4 bar absolute
Permissible operating temperature: 140°C

Site-specific media connections:
Electrical connection 380-400V, 50/60 Hz, 16A
Demineralized water: Gate valve 3/4" for hose connection (only for fixed connection, water can also be supplied from a storage tank)
Raw water: Gate valve 3/4" for hose connection (for exhaust steam condensation, device can also be operated without this)
Drain: Hose connection 3/4" max. 500 mm from floor or floor drain

Warranty: One year from delivery for material and manufacturing defects
3 years on heating elements
5 years on pressure vessels and closure mechanisms
Large product image